• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. FLUENCY PLUS ENDOVASCULAR STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. FLUENCY PLUS ENDOVASCULAR STENT GRAFT Back to Search Results
Model Number FEM14060
Medical Device Problem Codes Break (1069); Failure to Advance (2524); Migration (4003)
Health Effect - Clinical Codes Embolism/Embolus (4438); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/08/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records will be performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiry date: 05/2025).
 
Event or Problem Description
It was reported that during a stent graft placement procedure, the stent allegedly broke off when trying to deploy the stent inside the patient.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.Investigation summary: the sample was returned for evaluation.In the returned sample, the stent graft was completely deployed.The distal end of the outer catheter was damaged and the conical tip including marker band was missing.The images show the deployed stent graft in the vessel with the distal end constrained by the broken marker band.In a subsequent image the marker band is seen in the right lung.Provided images and returned sample could not confirm introducing resistance.Investigation leads to confirmation for break of the system tip and migration.Resistance was felt while crossing the venous graft segment, and the lesion was pre dilated.The investigation is closed with a confirmation for break of the system tip, and migration.A definite root cause for the reported event could not be determined based upon the available information.Labeling review: in reviewing the relevant labeling, it was found that the instructions for use sufficiently address the potential risks.The instructions for use state: 'if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device.' regarding preparation of the device the instructions for use state that 'prior to loading the endovascular system over a guide wire, both ports must be flushed with sterile saline (.).Flushing these lumens will also facilitate stent graft deployment.' regarding the anatomy of the placement site the instructions for use state: 'prior to stent graft deployment (.), ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of increased deployment forces and possible endovascular system failure.' regarding accessories the instructions for use states: 'prepare a stiff 0.035" guidewire per its instructions for use and advance the guidewire under fluoroscopy to the target location.The use of an appropriately sized introducer sheath is recommended.'; the packaging pictograms indicate an introducer size of 10f and a 0.035" guidewire.H10: d4 (expiry date: 05/2025).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during a stent graft placement procedure via subclavian vein, the tip of the sleeve allegedly broke off when trying to deploy the stent inside the patient.It was further reported that the device allegedly had resistance to passage of the pta balloon.Reportedly, the tip of the constraining sheath was embolized to the lung.The current status of the patient is unknown.
 
Event or Problem Description
It was reported that during a stent graft placement procedure via subclavian vein, the tip of the sleeve allegedly broke off when trying to deploy the stent inside the patient.It was further reported that the device allegedly had resistance to passage of the pta balloon.Reportedly, the tip of the constraining sheath was embolized to the lung.The current status of the patient is unknown.
 
Additional Manufacturer Narrative
H10: additional information was received and the file was reassessed for reportability and determined to be reportable as serious injury.H10: as the lot number for the device was provided, a review of the device history records will be performed.The device has been returned to the manufacturer for evaluation.However, a photo and images were provided for review.The investigation of the reported event is currently underway.H10: d4 (expiry date: 05/2025).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS ENDOVASCULAR STENT GRAFT
Common Device Name
ENDOVASCULAR STENT GRAFT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15922051
Report Number2020394-2022-00960
Device Sequence Number6698677
Product Code PFV
UDI-Device Identifier04049519008691
UDI-Public(01)04049519008691
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P130029
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFEM14060
Device Catalogue NumberFEM14060
Device Lot NumberANGS2332
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/23/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/09/2022
Supplement Date Received by Manufacturer12/06/2022
01/26/2023
Initial Report FDA Received Date12/06/2022
Supplement Report FDA Received Date01/02/2023
02/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age80 YR
Patient SexMale
Patient Weight86 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-