Catalog Number 381437
Medical Device Problem Code Retraction Problem (1536)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd insyte¿ autoguard¿ shielded iv catheter needle failed to retract after the insertion.The following information was provided by the initial reporter: "needle failed to retract from iv after insertion.".
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes, d10: returned to manufacturer on: 30-nov-2022.H6: investigation summary our quality engineer inspected the sample submitted for evaluation.Bd received one unsealed 20ga x 1.16in.Insyte autoguard.The unit was attempted to be retracted but failed.The button seemed to be stuck and would not click all the way.The device was microscopically analyzed for any damage to the grip and needle hub, but no damage was discovered.The coils also were inspected but no defects were discovered.The device was then forcefully retracted and then placed back into its original position, but then it was able to retract successfully without any issues.The reported issue was confirmed.Although adhesive was not present on the device, during manufacturing a station alignment may cause adhesive to be on the outside of the hub, causing partial or no retraction.Since the button was unable to be pressed originally, it can be assumed that there was some adhesive that was missed in the microscopic analysis and was then taken off after forcefully retracted.Or the very top coil near the button assembly was stuck, preventing the button from being pressed all the way.A device history record review showed no non-conformances associated with this issue during the production of this batch.H3 other text : see h10.
 
Event or Problem Description
It was reported that the bd insyte¿ autoguard¿ shielded iv catheter needle failed to retract after the insertion.The following information was provided by the initial reporter: "needle failed to retract from iv after insertion.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer (Section G)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15922053
Report Number1710034-2022-00750
Device Sequence Number6698679
Product Code FOZ
UDI-Device Identifier00382903814374
UDI-Public(01)00382903814374
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K201075
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number381437
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/10/2022
Supplement Date Received by Manufacturer01/18/2023
Initial Report FDA Received Date12/06/2022
Supplement Report FDA Received Date02/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-