Model Number VNMF3737C59TU
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Codes Stroke/CVA (1770); Insufficient Information (4580)
Date of Event 08/01/2021
Type of Reportable Event Serious Injury
Event or Problem Description
A valiant navion stent graft was implanted during the endovascular treatment of an aortic aneurysm.It was reported on an unknown date, follow up imaging identified an expanding abdominal aortic aneurysm with suspected occult type iii endoleak and impending loss of seal in the right iliac limb.Approximately 12 months post index procedure ultrasound guided needle placement right femoral artery, circumferential placement of endoanchors in the abdominal aortic endograft and placement of a right iliac limb extension was completed.It was reported approximately 12 months later the patient presented and was diagnosed with a r mca infarct and hospitalized.Treatment was given and the patient had a haemorrhagic conversion with ich, iv extension and sah.The patient was intubated but extubated for comfort care and the patient expired 2 days post hospitalisation.Per the physician the cause of the aneurysm expansion and type iii endoleak are undetermined.The cause of the stroke leading to death was not provided.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
B5; additional information received : it was reported there was no type iii endoleak.The endoleak was undetermined but suspected to be a occult ia or type ii.It was confirmed the endurant limb was implanted as the aneurysm had expanded and the limb was used as treatment for a possible type ii endoleak.The physician confirmed the stroke leading to the patient death had no relation to the endurant, heli-fx endoanchors or valiant navion stent graft.It was confirmed the endurant limb and endoanchors resolved the unknown endoleak and there are no complaints against these devices.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Core lab review of ct imaging from approximately 21 months post index procedure observed no endoleak, fracture, stent ring enlargement or stent ring migration.Core lab review of ct imaging from approximately 24 months post index procedure was not evaluable for endoleak, fracture or stent ring enlargement and no stent ring migration was observed.Core lab review of x-ray imaging from approximately 24 months post index procedure was not assessable for endoleak or stent ring enlargement, not evaluable for fracture and no stent ring migration was observed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALIANT NAVION
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key15922316
Report Number9612164-2022-04578
Device Sequence Number5218981
Product Code MIH
UDI-Device Identifier00763000100667
UDI-Public00763000100667
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P100040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/21/2021
Device Model NumberVNMF3737C59TU
Device Catalogue NumberVNMF3737C59TU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/08/2022
Supplement Date Received by Manufacturer01/18/2023
05/28/2024
Initial Report FDA Received Date12/06/2022
Supplement Report FDA Received Date01/26/2023
06/26/2024
Date Device Manufactured10/22/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Death;
Patient Age75 YR
Patient SexMale
Patient Weight105 KG
-
-