| Catalog Number |
07027583190 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The patient sample was received for investigation.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
The initial reporter stated they received questionable results for one patient sample tested with the elecsys troponin t hs on a cobas e 801 module (serial number (b)(4)).The troponin t hs results did not agree with the clinical examination and history of the patient and the customer suspects an interference-based discrepant high elecys troponin t hs result.The patient sample also generated questionable elecsys myoglobin results, which were not taken into consideration since the elecsys troponin t values were already considered implausible.This medwatch will apply to the elecsys myoglobin assay.Please refer to the medwatch with patient identifier.Patient identifier (b)(6) for information related to the elecsys troponin t hs assay.The sample initially resulted in a myoglobin value of 53.1 ng/ml.The sample was diluted 10 times and repeated, resulting in a value of <210 ng/ml, accompanied by a data flag.The sample was then repeated undiluted, resulting in a value of 21.0 ng/ml, accompanied by a data flag.The questionable results were reported outside of the laboratory.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon initial investigation of the patient sample, the customer's results could be reproduced.The patient sample resulted in a troponin t hs value of 275 pg/ml, a myoglobin value of 57 ng/ml, and a ck-mb value of 7.38 ng/ml.The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
A general reagent issue can be excluded.The investigation could not identify a product problem.The cause of the event could not be determined.
|
| |
|