• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS ALPHAVAC; ALPHAVAC MULTIPURPOSE MECHANICAL ASPIRATION 25F/C20

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS ALPHAVAC; ALPHAVAC MULTIPURPOSE MECHANICAL ASPIRATION 25F/C20 Back to Search Results
Model Number 25-302
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Non specific EKG/ECG Changes (1817); Hemorrhage/Blood Loss/Bleeding (1888); Decreased Respiratory Rate (2485)
Date of Event 11/11/2022
Type of Reportable Event Serious Injury
Event or Problem Description
This medwatch is not to report a device malfunction, but to report an adverse patient effect during a procedure in which an angiodynamics device was being used.There was no report of a device malfunction during the procedure.Physician gained access in r fv, swan-ganz introduced and navigated into pa, exchanged for lunderquist wire.Use of sequential dilators introduced at access site to accommodate 24f gore dry seal.Alphavac system was prepped within ifu guidance and introduced through 24f sheath over wire.Av outer sleeve with dilator was navigated over wire into pa.Dilator and wire exchanged for alphavac cannula and handle system.Noticed locking component on hub of alphavac sheath had abnormal resistance upon introduction of cannula.Exposed cannula into rpa, aspirated on 10cc option and flipping to 30cc option.Undesirable material was aspirated along with 100cc of blood.Physician used accessory port to introduce wire for navigation into lower rpa.Wire was taken out after gaining desired position and aspirated on 30cc.An additional 50cc of blood and material aspirated.Introduced multiuse diagnostic catheter through accessory port, did not achieve desired position.Took cannula out of outer sleeve of system, leaving the outer sheath in place.Introduced catheter through outer sheath into pa for contrast image.Exchanged catheter for wire and cannula, aspirating on 30cc after gaining desired positioning.Aspirated additional undesirable material and 50cc of blood.Removed alphavac system, patient stable with improved pa pressures.Within <3 min time span, physician began process of introducing filter when patient coughed blood and stats began declining.Patient taken to icu after intubation.Follow up on 11/14 confirmed patient stable; however, the patient still remains in cvicu".The patient had a prior procedure and additional comorbidities that attributed to the adverse patient event.The physician reported that the adverse patient outcome is unrelated to the device.
 
Additional Manufacturer Narrative
The device is not available to be returned to the manufacturer for evaluation.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALPHAVAC
Common Device Name
ALPHAVAC MULTIPURPOSE MECHANICAL ASPIRATION 25F/C20
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls, ny 12801 NY 12801
Manufacturer (Section G)
ANGIODYNAMICS
10 glens falls technical park
glens falls, ny 12801 NY 12801
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587805
MDR Report Key15923495
Report Number1319211-2022-00139
Device Sequence Number5209930
Product Code QEZ
UDI-Device Identifier15051684025966
UDI-PublicH787253020
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K211081
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number25-302
Device Catalogue Number25302
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/11/2022
Initial Report FDA Received Date12/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-