| Catalog Number |
010000937 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4) the device will not be returned for analysis.However; an investigation of the reported event is in progress.Once the investigation is completed, as supplemental medwatch will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during a replacement procedure the liner would not seat properly.Another liner was opened and seated without any issues.There were no consequences or impact to the patient.Attempts have been made and no further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Dhr was reviewed and no discrepancies related to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified there are indentations on the outside radius and on the lip of the liner.The locking feature of the device shows damage and shavings.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No additional event information to report at this time.
|
| |
|