Model Number DM3500
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/30/2022
Event Type  Injury  
Event Description
It was reported that the patient presented remotely via merlin.Net.Review of the transmission revealed that the implantable cardiac monitor (icm) failed to sense the r waves.The icm device was explanted and replaced with alternate icm device.The patient's condition was stable.
 
Manufacturer Narrative
The reported event of failure to sense was confirmed.The device was received with battery voltage above elective replacement indicator (eri) level.Analysis performed indicated that device failed sensing test, however, the device recovered during analysis and passed all tests.The cause of the sensing failure could not be determined.Longevity assessment was performed, and device was in the normal range of operation with appropriate remaining longevity.A device history record (dhr) review was performed and all required manufacturing processes and inspection steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONFIRM
Type of Device
IMPLANTABLE CARDIAC MONITOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key15923823
MDR Text Key304883729
Report Number2017865-2022-47820
Device Sequence Number1
Product Code MXC
UDI-Device Identifier05415067027320
UDI-Public05415067027320
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/04/2024
Device Model NumberDM3500
Device Catalogue NumberDM3500
Device Lot NumberP000155285
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/30/2022
Initial Date FDA Received12/06/2022
Supplement Dates Manufacturer Received03/27/2023
Supplement Dates FDA Received03/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/07/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
Patient Weight64 KG
-
-