Catalog Number 03.226.004
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/04/2022
Type of Reportable Event Malfunction
Event or Problem Description
Device report from synthes reports an event in colombia as follows: when using the screwdriver to place a screw, when turning it inside the screw head, it broke, it did not cause repercussions on the patient because it was done with another piece of screwdriver that was in the kit.It is marked with red tape.Concomitant device reported: unk - screws: trauma (part# unknown; lot# unknown; quantity: unknown) this report is for one (1) scrdriver shaft cann t8 w/colour-marking.This is report 1 of 1 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.Reporter is a j&j employee.Manufacturer site: postal code (b)(4).The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h4, h6: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable device history lot.Part #: 03.226.004.Lot #: 4l48064.The lot # 4l48064 does not exist in the system.Dhr review cannot be completed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D10: date of concomitant therapy is (b)(6) 2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCRDRIVER SHAFT CANN T8 W/COLOUR-MARKING
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES SELZACH
bohackerweg 5
selzach
SZ  
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key15924167
Report Number8030965-2022-10703
Device Sequence Number6689280
Product Code HXX
UDI-Device Identifier07611819194304
UDI-Public(01)07611819194304
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.226.004
Device Lot Number4L48064
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/10/2022
Supplement Date Received by Manufacturer12/21/2022
Initial Report FDA Received Date12/06/2022
Supplement Report FDA Received Date01/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - SCREWS: TRAUMA
Patient SexUnknown
-
-