• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 20 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 20 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number 20-800
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Capsular Contracture (1761)
Type of Reportable Event Serious Injury
Event or Problem Description
Patient reported left-side capsular contracture, baker grade unknown.Later, healthcare professional reported, capsular contracture baker grade iv.The device has been explanted.Treatment - capsulotomy.
 
Additional Manufacturer Narrative
The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iv.
 
Event or Problem Description
Patient reported left-side capsular contracture, baker grade unknown.Later, healthcare professional reported, capsular contracture baker grade iv.The device has been explanted.Treatment - capsulotomy.
 
Additional Manufacturer Narrative
Correction to h10 in mw2709736.Device evaluation: based on the device analysis grid, the assessments of the complaint are: capsular contracture: unable to confirm as it is a medical event not related to the device.Additional observations: crease fold observed.Wear abrasion observed.Observed, opening device on posterior side assessed as unidentified (tear) opening (shell thickness was within specification).No further actions are required as the device was implanted.
 
Additional Manufacturer Narrative
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Additional, changed, and/or corrected data: d1, d2, d4, d9, h4, h6.
 
Event or Problem Description
Patient reported left-side capsular contracture, baker grade unknown.Later, healthcare professional reported, capsular contracture baker grade iv.The device has been explanted.Treatment - capsulotomy.
 
Event or Problem Description
Patient reported left-side capsular contracture, baker grade unknown.Later, healthcare professional reported, capsular contracture baker grade iv.The device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 20 SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key15924185
Report Number9617229-2022-21883
Device Sequence Number14688151
Product Code FTR
UDI-Device Identifier10888628001800
UDI-Public10888628001800
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 08/17/2023
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/26/2020
Device Model Number20-800
Device Catalogue Number20-800
Device Lot Number2763091
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/18/2022
Supplement Date Received by Manufacturer01/03/2023
02/14/2023
07/27/2023
Initial Report FDA Received Date12/06/2022
Supplement Report FDA Received Date01/23/2023
03/08/2023
08/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/26/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight73 KG
-
-