• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VERATHON MEDICAL ULC GLIDESCOPE CORE 15-INCH MONITOR; LARYNGOSCOPE, RIGID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VERATHON MEDICAL ULC GLIDESCOPE CORE 15-INCH MONITOR; LARYNGOSCOPE, RIGID Back to Search Results
Model Number 0570-0404
Medical Device Problem Codes Display or Visual Feedback Problem (1184); No Visual Prompts/Feedback (4021)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/10/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that during a patient procedure, using a glidescope core 15-inch monitor, the image froze five (5) times during a difficult airway management.No delay in the procedure, use of a backup device, or harm to the patient was reported.
 
Additional Manufacturer Narrative
The reported glidescope core 15-inch monitor was returned to verathon for evaluation.A verathon technical service representative evaluated the reported video monitor but was unable to confirm the reported frozen image issue.When connecting the reported video monitor to known, good, test verathon equipment, the image produced on the video monitor was normal.The verathon technical service representatives was unable to confirm the reported frozen image issue when tested with verathon test equipment.The camera image quality test was performed and passed for the reported glidescope core 15-inch monitor.No additional information was provided regarding the verathon glidescope cable and camera used with the glidescope core 15-inch monitor during the reported incident.Upon completion of the evaluation the glidescope core 15-inch monitor was returned back to the customer.At this time no further investigation is required.Verathon will continue to monitor for any ongoing trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESCOPE CORE 15-INCH MONITOR
Common Device Name
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA  V5C 5A9
Manufacturer (Section G)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA   V5C 5A9
Manufacturer Contact
corey kasbohm
20001 n creek pkwy
bothell, WA 98011-8218
4256295760
MDR Report Key15924621
Report Number9615393-2022-00212
Device Sequence Number6317734
Product Code CCW
UDI-Device Identifier00879123006639
UDI-Public010087912300663911210416
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 11/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0570-0404
Device Catalogue Number0270-1033
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/10/2022
Initial Report FDA Received Date12/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/16/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-