• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE AUTOGUARD WINGED PNK 20GA X 1.16I; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE AUTOGUARD WINGED PNK 20GA X 1.16I; INTRAVASCULAR CATHETER Back to Search Results
Model Number 381534
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The initial reporter also notified the fda on 01-dec-2022.Medwatch report # mw (b)(4).A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported prior to use with bd insyte autoguard winged pnk 20ga x 1.16i foreign matter was found in 12 devices.There was no report of patient impact.The following information was provided by the initial reporter: black speckles in the plastic cap of the needle.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 12-dec-2022.Our quality engineer inspected the samples submitted for evaluation.Bd received two sealed 20g x 1.16in.Insyte autoguard winged units from lot number 2208315.A gross visual inspection found that both units had black specks of fm embedded in the needle cover.The reported issue was confirmed.This was physical/mechanical evidence to confirm and support a manufacturing process related issue for the reported defect.The material was determined to be burnt particulate resin (non-foreign).The black specks were most likely created as part of the molding process.Burnt imbedded resin specks result from material build up in the barrel/screw.A device history record review showed no non-conformances associated with this issue during the production of this batch.H3 other text : see h10.
 
Event or Problem Description
It was reported prior to use with bd insyte autoguard winged pnk 20ga x 1.16i foreign matter was found in 12 devices.There was no report of patient impact.The following information was provided by the initial reporter: black speckles in the plastic cap of the needle.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSYTE AUTOGUARD WINGED PNK 20GA X 1.16I
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer (Section G)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15925165
Report Number1710034-2022-00752
Device Sequence Number20415550
Product Code FOZ
UDI-Device Identifier00382903815340
UDI-Public(01)00382903815340
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K201075
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date07/31/2025
Device Model Number381534
Device Catalogue Number381534
Device Lot Number2208315
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/14/2022
Supplement Date Received by Manufacturer01/12/2023
Initial Report FDA Received Date12/06/2022
Supplement Report FDA Received Date01/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/04/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-