Model Number 03.820.111
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
Device report from depuy synthes reports an event in switzerland as follows: it was reported that intraoperatively on an unknown date in 2022 the device had a loose pin on one of the distractor arms.The distractor arm was fixed by putting a k-wire instead of the pin.The sterilization department found the missing/defective pin in the sterilization tray after the surgery.There was approximate five (5) minutes surgical delay.There was no patient impact.The procedure was successfully completed.During manufacturer's investigation of the returned device it was identified that the dowel pin, placed on the right retainer arm, was not fully seated, causing looseness of the retainer arm.This report is for one (1) vertebral body retainer this is report 1 of 1 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.A product investigation was conducted.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the dowel pin, placed on the right retainer arm, was not fully seated, causing looseness of the retainer arm.A dimensional inspection was unable to be performed due to not being applicable to the complaint condition.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed] as the observed condition of the vertebral body retainer, p/n: 03.820.111 would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.A device history record (dhr) review was conducted: part number: 03.820.111, lot number: 127p341 (wo# (b)(4)), manufacturing site: (b)(4), release to warehouse date: 04-may-2021.A review of the device history records was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERTEBRAL BODY RETAINER
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES TUTTLINGEN
unter hasslen 5
tuttlingen 78532
GM   78532
Manufacturer Contact
kate karberg
1302 wrights lane east
west chester, PA 19380
8472871282
MDR Report Key15925287
Report Number8030965-2022-10706
Device Sequence Number20102996
Product Code LXH
UDI-Device Identifier10705034731520
UDI-Public(01)10705034731520
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03.820.111
Device Catalogue Number03.820.111
Device Lot Number127P341
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/18/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/01/2022
Initial Report FDA Received Date12/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/04/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - GUIDE/COMPRESSION/K-WIRES
Patient SexUnknown
-
-