Device report from synthes reports an event in japan as follows: it was reported that on (b)(6) 2022, the patient underwent surgery with the products in question.During the surgery, the surgeon pointed out that the 2.0 lcp 1.5mm drill was difficult to attach to the locking hole of the plate and would come off immediately.Of the seven holes, the distal three holes could be attached, but the proximal four holes were difficult to attach.Another doctor who was assisting pointed out that the same phenomenon had been occurring for some time.The surgery was completed successfully with a 30-minute delay.The patient outcome was reported to be stable no further information is available.This report involves one 1.5mm threaded drill guide with depth gauge.This is report 2 of 2 for (b)(4).
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: complainant part is not expected to be returned for manufacturer review/investigation.Initial reporter facility name: (b)(6) hospital.Reporter is a j&j employee.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|