| Model Number |
120-340 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: textured implant exchange.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side implant removal from textured to smooth due to the concerns of the product.Healthcare professional later reported rupture.Device has been explanted and replaced.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side implant removal from textured to smooth due to the concerns of the product.Healthcare professional later reported rupture.Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: capsular contracture: unable to observe as it is a medical event not related to the device.Rupture: observed two openings assessed as an unidentified (tear) opening (shell thickness whiting to spec).Additional observations: red particles observed on the device.Additional, changed, and/or corrected data: h3, h6.
|
| |
|