Model Number MR17012601
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/27/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
We are currently waiting for the device to be returned for evaluation.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Picc leaking at site, removed, found to have a bleb ~ 1cm from hub and crack ~ 2 cm from hub.21g lumen affected, 23g lumen seems fine.Line inserted oct 24, never used for blood draws and no reported issues with high pressures/flushing line.
 
Additional Manufacturer Narrative
The 2.6f vascu picc was returned with a partially filled syringe attached to the end caps.The syringe and end caps were removed.The device hub is stained brown from approximately 0.5cm above the distal tip extending down approximately 1 cm of the lumen.The device was flushed, and a leak in the lumen was noted approximately 3 cm from the hub.Under magnification the lumen appears to have slightly ballooned around the 1.5 cm mark down to the 3 cm leak.A great deal of dried blood was noted just distal to the leak, suggesting the balloon and leak may have been caused by flushing against resistance.The device contract manufacturer conducted a review of the manufacture records for the lot number reported.Their investigation revealed the device was manufactured and inspected according to specification with no non-conformances or abnormalities.The manufacture process includes a 100% leak test performed as a last step in the process.This test would have detected any leaks, holes, or weak spots if it had existed at the time of manufacture.The instructions for use (ifu) contains the following warnings and precautions: * do not use infusion equipment which can exceed the working pressure of 1.0bar max/750mmhg (14.5 psi).*only use infusion equipment complying with standards, which do not exceed a shut-off pressure of 1.0bar.*bolus injections should be slow and must not exceed the maximum bolus pressure of 1.2bar/900mmhg (17.4 psi).*small syringes will generate excessive pressure and may damage the catheter.The use of 10cc or larger syringes are recommended.*do not flush against resistance.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.6F X 50CM DUAL LUMEN PICC
Common Device Name
VASCU-PICC
Manufacturer (Section D)
MEDICAL COMPONENTS, INC.
1499 delp drive
harleysville PA 19438
Manufacturer (Section G)
MEDICAL COMPONENTS, INC.
1499 delp drive
harleysville PA 19438
Manufacturer Contact
lynn winkler
1499 delp drive
harleysville, PA 19438
2152564201
MDR Report Key15926141
Report Number2518902-2022-00070
Device Sequence Number17449436
Product Code LJS
UDI-Device Identifier00884908085693
UDI-Public884908085693
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date03/01/2025
Device Model NumberMR17012601
Device Catalogue NumberMR17012601
Device Lot NumberMQHZ040 S2
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/06/2022
Supplement Date Received by Manufacturer12/06/2022
Initial Report FDA Received Date12/06/2022
Supplement Report FDA Received Date01/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/29/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-