Model Number TFAT00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Visual Impairment (2138)
Date of Event 08/23/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A sample device was not returned for analysis.Product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A nurse reported following an intraocular lens (iol) implant procedure, the patient could not adjust.The lens was explanted and replaced with monofocal lens in a secondary procedure.The clinical reason for the explant was mechanical failure to adapt.Additional information was requested, but further no information was available.
 
Additional Manufacturer Narrative
The lens was returned in a small glass container filled with a clear liquid and what appears to be a plastic cover.The optic was cut into pieces, typical of insertion and removal.Power and resolution testing could not be conducted due to the extensive optic damage.The file indicated the use of qualified associated products.Information was provided that in the surgeon's opinion, the lens did not cause or contribute to the event.In the surgeon's opinion the cause of the event was the patient could not adjust to the lens.The power and resolution of each lens is 100% evaluated during the manufacturing process to determine acceptability per model and diopter.Information was provided in the file that the multifocal company 21.0 diopter, add power +2.2 & 3.2 lens model was replaced with a non-company monofocal (20.0) due to the exchange of a multifocal lens to a monofocal lens, this may suggest that the replacement lens model was an improved choice for the patient's vision needs.No additional information has been provided at this time.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
Additional information was received stating there was no patient harm and patient was not hospitalized as a result of this event.There are two medical device reports associated with this patient.This report is associated with the right eye.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ PANOPTIX UV ABSORBING TRIFOCAL IOL
Common Device Name
LENS, MULTIFOCAL
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15926146
Report Number1119421-2022-02584
Device Sequence Number5225718
Product Code MFK
UDI-Device Identifier00380652429733
UDI-Public00380652429733
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P040020
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 12/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/16/2025
Device Model NumberTFAT00
Device Catalogue NumberTFAT00.210
Device Lot Number15050172
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/16/2022
Supplement Date Received by Manufacturer12/08/2022
Initial Report FDA Received Date12/06/2022
Supplement Report FDA Received Date12/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH III, CARTRIDGE D.; MONARCH III, INJECTOR.; ORA SYSTEM.; PROVISC.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
Patient SexFemale
-
-