| Model Number |
1012447-15 |
| Medical Device Problem Codes |
Material Separation (1562); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a lesion in the diagonal coronary artery.An nc trek balloon dilatation catheter (bdc) was being advanced into a guide catheter (gc), when strong resistance was met between the bdc and the gc, causing the shaft of the bdc to become separated inside of the guide catheter.The bdc and all separated pieces were simply removed without issue, no snare was required, and no material was left in the patient.There was no adverse patient effect and no clinically significant delay in the procedure.Another same size bdc was used to successfully complete the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is retained by hospital and not returning for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported difficulty advancing the device; however, the separation appears to be related to operational context.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Search Alerts/Recalls
|