| Model Number |
C501 |
| Medical Device Problem Codes |
Low Test Results (2458); Low Readings (2460)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/09/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter stated they received discrepant results for one patient urine sample tested with albt2 tina-quant albumin gen.2 (microalbumin) on a cobas 6000 c (501) module.No incorrect results were reported outside of the laboratory.The sample initially resulted in a microalbumin value of 324.8 mg/l.The sample was repeated, resulting in a value of < 0.7 mg/l with a data flag.The sample was also repeated on (b)(6) 2022, resulting in a value of < 0.8 mg/l with a data flag.The microalbumin reagent lot number was 65268801.The reagent expiration date was requested, but not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Na.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation could not identify a product problem.The cause of the event could not be determined.
|
| |
|