• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. LOGIC TIB INSERT IMPL CRC, SZ 3.5, 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. LOGIC TIB INSERT IMPL CRC, SZ 3.5, 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED Back to Search Results
Model Number N/A
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Insufficient Information (4580)
Date of Event 11/07/2022
Type of Reportable Event Serious Injury
Event or Problem Description
As reported, approximately two years post initial left side tka, the female patient had a revision due to poly wear.There was no breakage of device or surgical delay/prolongation.Patient was last known to be in stable condition following the event.Sales rep was unable to obtain x-rays & photos.The device is not available for evaluation due to hospital disposed the device.
 
Additional Manufacturer Narrative
Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
(h3) the revision reported may have been due to third body wear, patient- related conditions, instability, and/or malalignment between the femoral and tibial components, which lead to prosthesis wear.However, this cannot be confirmed because the tibial insert was not returned for evaluation.The most probable root cause associated with the reported event of ¿prosthesis wear¿ is associated with material damage to a surface, usually involving progressive loss or displacement of material, due to relative motion between that surface and a contacting substance or substances.
 
Additional Manufacturer Narrative
(h3) the revision reported may have been due to third body wear, patient- related conditions, instability, and/or malalignment between the femoral and tibial components, which lead to prosthesis wear.However, this cannot be confirmed because the tibial insert was not returned for evaluation.The most probable root cause associated with the reported event of ¿prosthesis wear¿ is associated with material damage to a surface, usually involving progressive loss or displacement of material, due to relative motion between that surface and a contacting substance or substances.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were corrected: d4: model number.D10 concomitant devices: (b)(4) 02-012-45-3535 lgc tibial fit tray cem sz 3.5f / 3.5tt.(b)(4) 02-010-04-0235 logic cr femoral por, left, sz 3.5.(b)(4) 200-02-38 three peg patella 38mm.The revision reported may have been due to third body wear, patient- related conditions, instability, and/or malalignment between the femoral and tibial components, which lead to prosthesis wear.However, this cannot be confirmed because the tibial insert was not returned for evaluation.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC TIB INSERT IMPL CRC, SZ 3.5, 9MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66 court
gainesville, FL 32653
3523771140
MDR Report Key15928223
Report Number1038671-2022-01573
Device Sequence Number6722826
Product Code JWH
UDI-Device Identifier10885862230447
UDI-Public10885862230447
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K123342
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 01/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/07/2025
Device Model NumberN/A
Device Catalogue Number02-012-51-3509
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/20/2022
Supplement Date Received by Manufacturer02/23/2023
02/23/2023
01/27/2025
Initial Report FDA Received Date12/06/2022
Supplement Report FDA Received Date03/15/2023
04/16/2024
01/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/08/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0021-2022
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LGC TIBIAL FIT TRAY CEM SZ 3.5F / 3.5TT; LOGIC CR FEMORAL POR, LEFT, SZ 3.5; THREE PEG PATELLA 38MM
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexFemale
-
-