| Model Number |
10839681 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported while using bd extension set smallbore trifuse 4 slide clamp leakage occurred.There was no report of patient impact.The following information was provided by the initial reporter: we have had a few reports of 10839681 leaking.
|
| |
|
Additional Manufacturer Narrative
|
|
Medical device expiration date: unknown.Device manufacture date: unknown.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd extension set smallbore trifuse 4 slide clamp leakage occurred.There was no report of patient impact.The following information was provided by the initial reporter: we have had a few reports of 10839681 leaking.
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 06-dec-2022.H6: investigation summary the customer reported extension sets leaking at connections sites and through lumens.The customer originally reported material 10839681, but returned 3 sets of mp5303-c.The complaint of leakage was tested by priming the samples, but was not verified.Device history record review for model mp5303-c lot number 22079022 was performed.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.
|
| |
|