Model Number 60005
Medical Device Problem Code Optical Problem (3001)
Health Effect - Clinical Code Unspecified Eye / Vision Problem (4471)
Date of Event 11/08/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device history record for the manufacturing lot was reviewed and there were no discrepancies or unusual findings.The treating physician reported that the patient felt her quality of vision was poor and spectacles does not improve; however, multiple repeat exams have shown no other abnormalities.Itrace shows some spherical aberration in the internal optics, and the treating physician suspects that she had some uv light exposure/leakage that may have irregularly polymerized the iol.The treating physician reports that during explantation, the lal was broken into many very small pieces and is unable to be returned for further evaluation.Manufacturer reference #: (b)(4).
 
Event or Problem Description
The site reported that a light adjustable lens (lal) was explanted due to patient dissatisfaction and suspected zone formation.Rxsight's first awareness of this event was on (b)(6) 2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGHT ADJUSTABLE LENS (LAL)
Common Device Name
LIGHT ADJUSTABLE LENS (LAL)
Manufacturer (Section D)
RXSIGHT, INC.
100 columbia
aliso viejo CA 92656
Manufacturer (Section G)
RXSIGHT, INC.
100 columbia
aliso viejo CA 92656
Manufacturer Contact
jim zhang
100 columbia
aliso viejo, CA 92656
9492736163
MDR Report Key15928468
Report Number3012712027-2022-00063
Device Sequence Number6563832
Product Code PZK
UDI-Device Identifier00818806020135
UDI-Public01008188060201351119121117221130
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P160055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2022
Device Model Number60005
Device Catalogue Number60005
Device Lot NumberL02-001471
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/09/2022
Initial Report FDA Received Date12/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/11/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexFemale
-
-