MAUDE Adverse Event Report: MEDTRONIC MEXICO S. DE R.L. DE CV ACHIEVE MAPPING CATHETER - 15 MM; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO S. DE R.L. DE CV ACHIEVE MAPPING CATHETER - 15 MM; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECT Back to Search Results
Model Number 990063-015
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/06/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a cryo ablation procedure, noise was heard coming from two electrodes on the mapping catheter.The mapping catheter cable was replaced without resolution.Subsequently, the mapping catheter was replaced which resolved the issue. the case was completed with cryo. no patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the 990063-015 mapping catheter with lot 224571488 was returned and analyzed.Visual inspection of the shaft segment area showed that the shaft was kinked approximately 5.5 inches from the lemo connector.Functional testing was then performed.The mapping catheter was connected to the diagnostic computer (prucka) using the test cable.The channel pairs 2-3 and 3-4 were displaying noise.The rest of the channels were normal.Dissection of the suspected electrode related to the noise showed the electrode wire was connected properly to the electrode and it is intact with no issues.Dissection of lemo connector assembly showed broken electrical wire inside lemo connector.In conclusion, the mapping catheter tested out of specification and failed the returned product analysis due to a shaft kink.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACHIEVE MAPPING CATHETER - 15 MM
Common Device Name
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECT
Manufacturer (Section D)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX  22210
Manufacturer (Section G)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX   22210
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key15928719
Report Number9617601-2022-00188
Device Sequence Number6563839
Product Code DRF
UDI-Device Identifier00643169467378
UDI-Public00643169467378
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K102588
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/27/2024
Device Model Number990063-015
Device Catalogue Number990063-015
Device Lot Number224571488
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/06/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/10/2022
Initial Report FDA Received Date12/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/27/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age68 YR
Patient SexMale
Patient Weight103 KG
-
-