| Model Number |
PCB00 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Information unknown/ not provided.Per regulation eu 2016/679 (general data protection regulation), patient identifiers were not collected or recorded and therefore are not available.If implanted, give date: not applicable, as lens was not implanted.If explanted, give date: not applicable, as lens was not implanted.Telephone number: (b)(6).Device evaluation: the preloaded unit was not returned for evaluation.Therefore, a failure analysis of the complaint device could not be performed.Manufacturing record evaluation: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search revealed that no additional complaints for this order number have been received.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the physician noticed a crack in the preloaded lens.The issue was identified during handling and prior to insertion with no patient contact.Through follow-up it was learned that there were no issues or complications for the patient.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: section d9: device available for evaluation? yes.Section d9: returned to manufacturer on: 09 feb 2023.Corrected data: upon further review it was noted that "g4" field which should have been populated with "no" was inadvertently left blank on the initial mdr; therefore, the information has been corrected in this supplemental mdr report and the following field has been updated accordingly: section g4: combination product: no.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection under magnification revealed that the complaint handpiece was received with the lens stuck inside of the cartridge.The cartridge could be observed to be bulging around the lens, and a crack on the side of the cartridge could be observed as well.The handpiece was disassembled and the assembly was inspected.No issues that could cause or contribute to the complaint issue were identified.The lens was removed from the handpiece and cleaned, revealing scratches and damage consistent with a lens that was stuck inside of a cartridge.Conclusion: the complaint issue of lens damaged was identified during the product evaluation; however, based on the complaint investigation results the complaint issue could not be confirmed to be related to the manufacturing or design process.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Therefore, there is no indication of a product deficiency or product malfunction.Therefore, there is no indication of a product deficiency or product malfunction.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|