Model Number PM2272
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Wound Dehiscence (1154); Unspecified Infection (1930)
Event Type  Injury  
Event Description
Related manufacturer reference number: 2017865-2022-47953, 2017865-2022-47954.It was reported that the patient presented to the clinic with discharge at the device site.It was noted that the incision was not healing.The pacemaker, right atrial lead and right ventricular lead were explanted due to pocket infection.The patient was in stable condition throughout the procedure.
 
Manufacturer Narrative
A device history record (dhr) review was performed and all required manufacturing processes and inspection steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.Review of the sterilization records confirmed normal sterilization cycles for the products.The product was returned and visual inspection was normal.Interrogation of the device revealed it was above elective replacement indicator (eri) when received.The cause of infection could not be traced to the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURITY MRI
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key15929375
MDR Text Key304955297
Report Number2017865-2022-47952
Device Sequence Number1
Product Code LWP
UDI-Device Identifier05414734509589
UDI-Public05414734509589
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/06/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/29/2024
Device Model NumberPM2272
Device Catalogue NumberPM2272
Device Lot NumberP000153251
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/27/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age63 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-