• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD CARIBE LTD. BD PERISAFE¿ EPIDURAL MINI-KIT; ANESTHESIA CONDUCTION KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD CARIBE LTD. BD PERISAFE¿ EPIDURAL MINI-KIT; ANESTHESIA CONDUCTION KIT Back to Search Results
Catalog Number 400273
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/11/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported while using (brand name) leakage occurred.There was no report of patient impact.The following information was provided by the initial reporter: leak during injection pressure with a 5ml syringe.
 
Additional Manufacturer Narrative
Date of event is unknown; awareness date has been used for this field.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: b5: describe event or problem: it was reported while using bd perisafe¿ epidural mini-kit leakage occurred.There was no report of patient impact.The following information was provided by the initial reporter: leak during injection pressure with a 5ml syringe.D10: device available for eval yes.D10: returned to manufacturer on: 16-dec-2022.One photo sample and one physical sample was provided to our quality team for investigation.The sample displays the material information as reported, however, there is no defect that can be identified within the sample, therefore we cannot verify the reported issue a device history review was performed for the reported lot 0036486 and no deviations or non conformances were identified during the manufacturing process that could have contributed to this issue.Based on the available information, we cannot identify a root cause at this time.
 
Event or Problem Description
It was reported while using bd perisafe¿ epidural mini-kit leakage occurred.There was no report of patient impact.The following information was provided by the initial reporter: leak during injection pressure with a 5ml syringe.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PERISAFE¿ EPIDURAL MINI-KIT
Common Device Name
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
BD CARIBE LTD.
road 31
k.m. 24.3
juncos
Manufacturer (Section G)
BD CARIBE LTD.
road 31
k.m. 24.3
juncos
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15929826
Report Number2618282-2022-00088
Device Sequence Number14548892
Product Code CAZ
UDI-Device Identifier50382904002737
UDI-Public00382904002732
Combination Product (Y/N)N
Initial Reporter CountryTS
PMA/510(K) Number
K953790
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date02/28/2025
Device Catalogue Number400273
Device Lot Number0036486
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/11/2022
Supplement Date Received by Manufacturer12/19/2022
Initial Report FDA Received Date12/07/2022
Supplement Report FDA Received Date01/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/13/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-