STRYKER ORTHOPAEDICS-MAHWAH TRIDENT HEMISPHERICAL SOLID BACK SHELL; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 500-11-48D
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Joint Dislocation (2374); Insufficient Information (4580)
Date of Event 11/11/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Patient booked for hip revision surgery (planned surgery).Surgeon planned for a replacement of femoral head.Photos of explanted devices indicate a ceramic head, alumina insert and acetabular shell were revised.
 
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Additional Manufacturer Narrative
An event regarding revision involving a trident shell was reported.The event was not confirmed.Method & results: product evaluation and results: visual inspection: the shell was returned with the liner still inside.Damage is visible consistent with explantation damage.Bone ingrowth is visible on the shell.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to dislocation following a fall.The reason for revision of the shell was not reported.The exact cause of the event could not be determined because insufficient information was provided.Further information such as pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.It is likely that patient trauma contributed to the device failure as it was reported that the patient fell prior to the dislocation.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Patient booked for hip revision surgery (planned surgery).Surgeon planned for a replacement of femoral head.Photos of explanted devices indicate a ceramic head, alumina insert and acetabular shell were revised.Patient was revised due to dislocation after falling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT HEMISPHERICAL SOLID BACK SHELL
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
joann ripoli
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15930133
Report Number0002249697-2022-01781
Device Sequence Number11861467
Product Code LPH
UDI-Device Identifier04546540431844
UDI-Public04546540431844
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K143085
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2006
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2011
Device Model Number500-11-48D
Device Catalogue Number500-11-48D
Device Lot Number18809201
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/11/2022
Supplement Date Received by Manufacturer04/03/2023
Initial Report FDA Received Date12/07/2022
Supplement Report FDA Received Date04/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/14/2006
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age72 YR
Patient SexFemale
Patient Weight73 KG
-
-