MAUDE Adverse Event Report: BIOMET UK LTD. OXF TWIN-PEG CMNTD FEM LG PMA; OXFORD CEMENTED FEMORALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET UK LTD. OXF TWIN-PEG CMNTD FEM LG PMA; OXFORD CEMENTED FEMORALS Back to Search Results
Model Number N/A
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Metal Related Pathology (4530)
Date of Event 11/09/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Medical devices: oxf anat brg lt lg size 4 pma; item# 159555; lot# 948510; oxf uni tib tray sz c lm pma; item# 154722; lot# 813810.Multiple mdr reports were filed for this event, please see associated reports: 3002806535 - 2022 - 00485; 3002806535 - 2022 - 00487.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
It was reported that patient underwent a partial knee arthroplasty and approximately 8 years later a revision surgery was performed due to metallosis.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.No products were returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Devices are used for treatment.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: "significant findings include lateral subluxation of the left tibia with respect to the femur and disrupted alignment of the medial compartment, suspected to be related to displaced bearing component which would have resided in the medial compartment but appears displaced superiorly.".A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXF TWIN-PEG CMNTD FEM LG PMA
Common Device Name
OXFORD CEMENTED FEMORALS
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key15930625
Report Number3002806535-2022-00486
Device Sequence Number6563390
Product Code NRA
UDI-Device Identifier05019279438259
UDI-Public(01)05019279438259(17)240311(10)816300
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
P010014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/10/2024
Device Model NumberN/A
Device Catalogue Number161470
Device Lot Number816300
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/09/2022
Supplement Date Received by Manufacturer01/18/2023
Initial Report FDA Received Date12/07/2022
Supplement Report FDA Received Date01/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/11/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-