• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VASORUM LTD. CELT ACD VASCULAR CLOSURE DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VASORUM LTD. CELT ACD VASCULAR CLOSURE DEVICE Back to Search Results
Model Number KCLT-05
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/02/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that physician attempted to use a 5f celt to close a 5f arterial access puncture.The physican noted that when he attempted to deploy the distal wing, the handle did not fully pop back.An ultrasound was performed which revealed that the distal wings did not fully open.The sheath and the delivery system with the implant attached were removed from the patient and manual compression was applied to achieve hemostasis.The patient had no complications and was discharged the same day.No implant remained in the patient.
 
Additional Manufacturer Narrative
During processing of this complaint, attempts were made to obtain complete event, patient and device information.A search of the complaints files didn't find any other complaint with the associated lot number.The device with the implant attached has been returned for detailed analysis.This report is the initial report being submitted by vasorum.A follow up report will be submitted as soon as further information on this event becomes available.
 
Event or Problem Description
It was reported that physician attempted to use a 5f celt to close a 5f arterial access puncture.The physican noted that when he attempted to deploy the distal wing, the handle did not fully pop back.An ultrasound was performed which revealed that the distal wings did not fully open.The sheath and the delivery system with the implant attached were removed from the patient and manual compression was applied to achieve hemostasis.The patient had no complications and was discharged the same day.No implant remained in the patient.
 
Additional Manufacturer Narrative
During processing of this complaint, attempts were made to obtain complete event, patient and device information.A search of the complaints files didn't find any other complaint with the associated lot number.The device with the implant attached has been was returned for detailed analysis.A damaged component was identified.This may have led to the failure of the implant to deploy.However, a definitive conclusion as to the cause of the damage to the component was not identified.All available information has been placed on file in the customer complaint files of vasorum ltd for appropriate tracking, trending and follow-up.This report is the final report being submitted by vasorum.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELT ACD VASCULAR CLOSURE DEVICE
Common Device Name
VASCULAR CLOSURE DEVICE
Manufacturer (Section D)
VASORUM LTD.
2012 orchard avenue
citywest business campus
dublin, D24WP W6
EI  D24WPW6
Manufacturer (Section G)
VASORUM LTD.
2012 orchard avenue
citywest business campus
dublin, D24W9 W6
EI   D24W9W6
Manufacturer Contact
2012 orchard avenue
citywest business campus
dublin, D24W9-W6
MDR Report Key15930981
Report Number3009984513-2022-00013
Device Sequence Number4347419
Product Code MGB
UDI-Device Identifier05391530280006
UDI-Public(01)05391530280006(17)250830(10)919603
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P150006
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/30/2025
Device Model NumberKCLT-05
Device Lot Number919603
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/02/2022
Supplement Date Received by Manufacturer11/02/2022
Initial Report FDA Received Date12/07/2022
Supplement Report FDA Received Date02/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-