MAUDE Adverse Event Report: STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP NEUT. ELECT, SELF-ADHESIVE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP NEUT. ELECT, SELF-ADHESIVE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number 0406-650-205
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Partial thickness (Second Degree) Burn (2694)
Date of Event 11/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Discarded by customer.
 
Event or Problem Description
It was reported that the patient received a burn at the grounding pad site, it was noted that grounding pad gel had dried.The procedure was completed successfully with no delays or medical intervention.A topical cream was used to treat the burn and prevent infection.
 
Event or Problem Description
It was reported that the patient received a burn at the grounding pad site, it was noted that grounding pad gel had dried.The procedure was completed successfully with no delays or medical intervention.A topical cream was used to treat the burn and prevent infection.
 
Additional Manufacturer Narrative
Follow-up report submitted to document device evaluation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUT. ELECT, SELF-ADHESIVE
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP
1941 stryker way
portage MI 49002
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
colette chung
1941 stryker way
portage, MI 49002
2693237700
MDR Report Key15931122
Report Number3015967359-2022-02400
Device Sequence Number6563406
Product Code GEI
UDI-Device Identifier35060842920023
UDI-Public35060842920023
Combination Product (Y/N)N
Initial Reporter StateSD
Initial Reporter CountryUS
PMA/510(K) Number
K900332
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0406-650-205
Device Lot Number202401BB
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/08/2022
Supplement Date Received by Manufacturer11/08/2022
Initial Report FDA Received Date12/07/2022
Supplement Report FDA Received Date03/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-