• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. FIBERTAK DOUBLE LOADED W/ 1.3 SUTURETAPE; FASTENER, FIXATION, NON-DEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. FIBERTAK DOUBLE LOADED W/ 1.3 SUTURETAPE; FASTENER, FIXATION, NON-DEGRADABLE, SOFT TISSUE Back to Search Results
Model Number FIBERTAK DOUBLE LOADED W/ 1.3 SUTURETAPE
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 11/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that during a shoulder surgery the tip of the device broke off and remained in the bone of the patient.The surgery time has not increased due to the error.The surgery was finished successfully with the same device.It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
The complaint is confirmed.One unpackaged ar-3602-2 serial/batch number 14973717 was received for investigation.Visual evaluation found that the fibertak shaft tip was bent and broken off.The fragments generated during the event were not returned for analysis.More in-depth evaluation shows that the device's shaft was bent.The root cause of the reported failure mode is undetermined.However, the most probable cause is applying excessive mechanical forces upon insertion/use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIBERTAK DOUBLE LOADED W/ 1.3 SUTURETAPE
Common Device Name
FASTENER, FIXATION, NON-DEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key15931161
Report Number1220246-2022-05899
Device Sequence Number20469056
Product Code MBI
UDI-Device Identifier00888867206953
UDI-Public00888867206953
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K181769
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 02/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFIBERTAK DOUBLE LOADED W/ 1.3 SUTURETAPE
Device Catalogue NumberAR-3602-2
Device Lot Number14973717
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/23/2022
Supplement Date Received by Manufacturer11/23/2022
Initial Report FDA Received Date12/07/2022
Supplement Report FDA Received Date02/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-