| Model Number |
434900 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/09/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During treatment of a type iii perforation in the rca, the pk papyrus covered stent system could not pass the pre-dilated severely calcified lesion and was withdrawn.The intervention was completed with another pk papyrus, and the perforation was successfully sealed.
|
| |
|
Additional Manufacturer Narrative
|
|
The affected pk papyrus stent system was not returned to biotronik and could therefore not be subjected to a technical investigation.Images of the case such as angiographies could also not be obtained.The provided clinical information (pk papyrus device registration form, cathlab report) and the product release documentation was reviewed to establish whether a device deficiency contributed to this event.Review of the product release documentation confirmed that the device was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.Based on the provided documentation and the conducted review no manufacturing related root cause could be identified.The treating physician rated the outcome as no device- and no procedure-related complication.Pk papyrus is indicated for the treatment of acute perforations of native coronary arteries and coronary bypass grafts in vessels 2.5 to 5.0 mm in diameter.As such, this device is typically used in emergent response to a life-threatening adverse procedural event unrelated to use of the pk papyrus.
|
| |
|
Search Alerts/Recalls
|