|
MEDOS INTERNATIONAL SÃ RL CH EXPEDIUM SPINE SYSTEM SI POLYAXIAL EXTEND TAB 5.5 X 5 X 30MM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
|
Back to Search Results |
|
| Model Number |
179732530 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: additional device product codes: kwq, mnh, osh, kwp, mni.Complainant part is not expected to be returned for manufacturer review/investigation.Reporter is a j&j employee.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from synthes reports an event in japan as follows: it was reported that this was a spinal fusion (th8 to l3) for the thoracolumbar vertebrae on (b)(6) 2022.During insertion of the rod, back-out of the six screws in question occurred at l1 to l3.It seemed that bone quality was declining because of the patient age.Screws were reinserted by increasing the screw size or changing the trajectory of the insertion hole.The surgery was completed successfully within 30 minutes delay.The patient outcome was reported to be stable.No further information is available.This report involves one expedium spine system si polyaxial extend tab 5.5 x 5 x 30mm.This is report 2 of 6 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: h4 a manufacturing record evaluation was performed for the finished device.Product code: 179732530.Lot number: 109057.It was electronically reviewed and no non-conformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: 15/11/2016.Qty: (b)(4).Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|