| Model Number |
2307-93-000 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the tip of the driver is worn out and not catching screws.There was no surgical delay.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d9 if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: h3.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4) this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: the device associated with this report was returned to depuy synthes for evaluation.Visual examination found wear and rounding of the driver's hex tip, confirming the reported allegation.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|