• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. TIBIA CEMENTED 5 DEGREE STEMMED LEFT SIZE D; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. TIBIA CEMENTED 5 DEGREE STEMMED LEFT SIZE D; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Codes Loss of or Failure to Bond (1068); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 11/18/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated report: 3007963827-2022-00313.Concomitant medical products: femur cemented (ps) standard left size 7, item#: 42500606201, lot#: 64824490; clearmix triple mix, item#: 414703, lot#: 31054; all-poly patella cemented 29 mm diameter, item#: 42540200029, lot#: 64835013; articular surface fixed bearing (ps) left 10 mm height, item#: 42512400510, lot#: 64832247; persona cem fem/cem tib/v, item#: 98000241520, lot#: 65315038.The product has been returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that a patient was revised due to loosening of the tibial and femoral components.Attempts to obtain additional information have been made; however, no more information is available at this time.
 
Additional Manufacturer Narrative
This follow-up is being submitted to relay additional information.
 
Event or Problem Description
It was reported that a patient was revised one year and five months post implantation due to loosening of the tibial and femoral components.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
Visual examination of the returned product identified signs of use.No bone cement was noted on both the femur and tibial plate.Radiographs were provided and reviewed by a healthcare professional.Review of the available records identified that due to poor imaging loosening cannot be confirmed.Device history record (dhr) was reviewed for deviations and/ or anomalies with no deviations / anomalies identified.A definitive root cause cannot be determined.Per package insert: 87-6204-022-23: loosening is a known risk of the system.If any further information is which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIA CEMENTED 5 DEGREE STEMMED LEFT SIZE D
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15931558
Report Number3007963827-2022-00312
Device Sequence Number11879039
Product Code MBH
UDI-Device Identifier00889024471030
UDI-Public(01)00889024471030(17)310112(10)64901868
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K172524
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42532006701
Device Lot Number64901868
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/18/2022
Supplement Date Received by Manufacturer12/03/2022
03/13/2023
Initial Report FDA Received Date12/07/2022
Supplement Report FDA Received Date12/21/2022
03/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age61 YR
Patient SexMale
-
-