• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEUWAVE MEDICAL, INC. NEUWAVE LK XT PROBE 15CM 15GA; SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEUWAVE MEDICAL, INC. NEUWAVE LK XT PROBE 15CM 15GA; SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES Back to Search Results
Catalog Number LK15XT
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an ablation procedure the tip of the probe broke.It was easily retrieved and a new probe was used to complete the case with no patient consequences.
 
Additional Manufacturer Narrative
(b)(4).Event year only reported: 2022.Batch #: unknown.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The lot/batch was not provided; therefore, the manufacturing records evaluation could not be performed.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 1/9/2023.Investigation summary: call home did not reveal any issues or errors.The probe returned and the broken probe tip was not verified.However, two bends were seen on the cannula shaft.No further information was available.The root cause of the bends is unknown.A search of nonconformities (ncs) was performed for lot ml20105726.There were no observed nonconformities for this lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUWAVE LK XT PROBE 15CM 15GA
Common Device Name
SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES
Manufacturer (Section D)
NEUWAVE MEDICAL, INC.
3529 anderson st
madison WI 53704
Manufacturer (Section G)
NEUWAVE MEDICAL, INC.
3529 anderson st
madison 53704
Manufacturer Contact
kate karberg
3529 anderson st
madison, WI 53704
3035526892
MDR Report Key15932363
Report Number3008769756-2022-00058
Device Sequence Number4383911
Product Code NEY
UDI-Device Identifier00853719006364
UDI-Public853719006364
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K160936
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberLK15XT
Device Lot NumberML20105726
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/15/2022
Supplement Date Received by Manufacturer12/19/2022
Initial Report FDA Received Date12/07/2022
Supplement Report FDA Received Date01/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-