• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 NIAGARA SLIM-CATH DIALYSIS CATHETER KIT 12F X 15CM (STRAIGHT); CATHETER, HEMODIALYSIS, NON-IMPLANTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 NIAGARA SLIM-CATH DIALYSIS CATHETER KIT 12F X 15CM (STRAIGHT); CATHETER, HEMODIALYSIS, NON-IMPLANTED Back to Search Results
Model Number N/A
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturer has received the sample and is pending evaluation.Results are expected soon.A lot history review (lhr) of refr3767 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported that there were black foreign matters with the guide wire.No other information was provided.
 
Event or Problem Description
It was reported that there were black foreign matters with the guide wire.No other information was provided.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, complaint and batch history, applicable previous investigation(s), labeling, applicable manufacturing records, sample analysis and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a foreign matter on the device is inconclusive due to poor sample condition.One photo sample of a guidewire wire being held was provided for evaluation.The guidewire was observed to contain a curve near the distal end.A darkened spot on the guidewire appeared to be present; however, it could not be clearly determined if the dark spot was discoloration or a residual material.The open kit and catheter contents are visible in the background.The lack of a returned sample did not allow for an inspection of the discolored region; therefore, the source and type of material could not be identified.Based on the information provided, possible contributing factors include contamination of kit component during handling or use.A review of the manufacturing records did not reveal any evidence to suggest that a manufacturing related root cause contributed to the reported event.The condition of the component prior to kit opening and handling is also unable to be determined; therefore, the complaint could not be confirmed and remains inconclusive at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIAGARA SLIM-CATH DIALYSIS CATHETER KIT 12F X 15CM (STRAIGHT)
Common Device Name
CATHETER, HEMODIALYSIS, NON-IMPLANTED
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kayla olsen
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key15932624
Report Number3006260740-2022-05607
Device Sequence Number19929435
Product Code MPB
UDI-Device Identifier00801741045356
UDI-Public(01)00801741045356
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K010778
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2023
Device Model NumberN/A
Device Catalogue Number5553150
Device Lot NumberREFR3767
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/18/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/16/2022
Supplement Date Received by Manufacturer01/18/2023
Initial Report FDA Received Date12/07/2022
Supplement Report FDA Received Date01/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-