| Model Number |
8900-0211-01 |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/09/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to cardiovert a patient (age & gender unknown), the associated defibrillator was unable to detect the attached electrode pads.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The actual electrode pads were not returned to zoll medical corporation for evaluation.Instead, a lot number of the pads was provided.An investigation was conducted on retained samples of onestep resuscitation electrode, lot 3522 for the customer's report.Evaluation of the electrodes did not reveal any irregularities that would have contributed to the reported event.Based on the evaluation of the retained samples it was concluded that the samples were assembled according to specification without duplicating the report.Analysis of reports of this type has not identified an increase in trend.
|
| |
|