• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA GMK-HINGE FIXED TIBIAL TRAY SIZE 5 R; KNEE TIBIA REVISION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA GMK-HINGE FIXED TIBIAL TRAY SIZE 5 R; KNEE TIBIA REVISION Back to Search Results
Model Number 02.09.4005R
Medical Device Problem Code Osseointegration Problem (3003)
Health Effect - Clinical Code Pain (1994)
Date of Event 11/18/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Clinical evaluation performed by medical affairs department.One year after a complex case of revision tka the cemented components become loose.The low-quality pictures included in the report show very low quality bones, possibly affected by some sort of osteomalacia, not mentioned in the report - a situation that certainly makes the fixation of a large constrained prosthesis very challenging.We can only hypothesize that the low quality of the bone made fixation impossible, because the implant position, albeit very challenging, appears correct.Batch review performed on 21 november 2022: lot 2101290: (b)(4) items manufactured and released on 09-june-2021.Expiration date: 2026-05-25.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event in the period of review.Other component involved: gmk-hinge 02.09.2604r femoral component size 4 r (k130299) lot 1904680: (b)(4) items manufactured and released on 25-sept-2019.Expiration date: 2024-09-07.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event in the period of review.
 
Event or Problem Description
Revision surgery at about 1 year and 1 month post primary due to the femur and tibial tray loosening.All components revised successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GMK-HINGE FIXED TIBIAL TRAY SIZE 5 R
Common Device Name
KNEE TIBIA REVISION
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ   6874
Manufacturer Contact
stefano baj
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key15932990
Report Number3005180920-2022-00923
Device Sequence Number6563227
Product Code KRO
UDI-Device Identifier07630030825606
UDI-Public07630030825606
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K130299
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/25/2026
Device Model Number02.09.4005R
Device Catalogue Number02.09.4005R
Device Lot Number2101290
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/18/2022
Initial Report FDA Received Date12/07/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient SexMale
-
-