Model Number M001468080
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 11/02/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a complex intracerebral thrombectomy procedure , the third probing maneuver with a guidewire (subject device) resulted in the fracture of the wire approximately 4-5 cm of the distal part.This was currently located in the middle cerebral artery.A retrieval of the fractured guidewire through a loop and it was completely removed without complications.No other information was provided.
 
Additional Manufacturer Narrative
The subject device is unavailable to manufacturer.
 
Additional Manufacturer Narrative
D4 expiration date - added.H4 manufacturing date ¿updated.Based on the results of the dhr (device history record) review, there is no indication that the device, labeling or packaging failed to meet its specifications when released.The subject device was not returned; therefore, a visual inspection as well as a functional evaluation could not be performed.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.While there are a number of potential causes for the reported issue, because review of available information failed to identify a definitive cause and the device was not returned for analysis, an assignable cause of undeterminable was assigned to this complaint.
 
Event or Problem Description
It was reported that during a complex intracerebral thrombectomy procedure , the third probing maneuver with a guidewire (subject device) resulted in the fracture of the wire approximately 4-5 cm of the distal part.This was currently located in the middle cerebral artery.A retrieval of the fractured guidewire through a loop and it was completely removed without complications.No other information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANSEND EX 14 PLATINUM 205CM
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC COSTA RICA
coyol facility, 2546 first st.
propark free zone
alajuela 20101
CS  20101
Manufacturer (Section G)
BOSTON SCIENTIFIC COSTA RICA
coyol facility, 2546 first st.
propark free zone
alajuela 20101
CS   20101
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key15933029
Report Number3008853977-2022-00057
Device Sequence Number15134125
Product Code DQX
UDI-Device Identifier08714729326267
UDI-Public08714729326267
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K971254
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/23/2023
Device Model NumberM001468080
Device Catalogue NumberM001468080
Device Lot Number28606074
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/30/2022
Supplement Date Received by Manufacturer01/31/2023
Initial Report FDA Received Date12/07/2022
Supplement Report FDA Received Date01/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age30 YR
Patient SexFemale
-
-