| Model Number |
VASOVIEW 7 XB BISECTOR (OUS) |
| Medical Device Problem Code |
Failure to Cut (2587)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The hospital reported that during an endoscopic vein harvesting procedure the vasoview 7 xb bisector (ous) was unable to cut and coag the patient¿s saphenous vein at cut setting of 30 and 40.The jaws are fully closed during insertion.There was no resistance.The procedure was completed with a new vasoview 7xb and a new karl storz autocon ii 200 esu.No delay on the procedural.No harm to patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Trackwise id (b)(4).The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.
|
| |
|
Additional Manufacturer Narrative
|
|
Trackwise#: (b)(4).
|
| |
|
Event or Problem Description
|
|
The hospital reported that during an endoscopic vein harvesting procedure the vasoview 7 xb bisector (ous) was unable to cut and coag the patient¿s saphenous vein at cut setting of 30 and 40.It failed to cut the tissue on activation.The jaws are fully closed during insertion.There was no resistance.User suspected it could be a faulty esu too.The procedure was completed with a new vasoview 7xb and a new karl storz autocon ii 200 esu.No delay on the procedural.No harm to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Trackwise#: (b)(4).The device was returned to the factory for evaluation on 11/28/2022.An investigation was conducted on 12/01/2022.A visual inspection was conducted.Signs of clinical use and no evidence of blood was observed.There were no visual defects observed on the bipolar blades or the cutter blade.A mechanical evaluation was conducted.The blue toggle was able to manipulated to retract and extend the cutter blade.An electrical evaluation was conducted.The device was evaluated for electrical function.A pre-cautery test was performed and repeated 10 times over a 10 minute period according to the procedure in the instructions for use (cv000001599).The device passed the pre-cautery test with a reference generator (recommended setting at 18) and reference bipolar cord.An activation test was performed and the device activated repeatedly with no failure observed.The bipolar cord connection was manipulated during activation.The device remained active and no sparks was observed.The device produced steam and the saline was observed to ¿boil¿ on the test gauze each time and no sparks were observed.Based on the returned condition of the device as well as the evaluation results, the reported failure "failure to cut" was not confirmed.The lot # 25163305 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|