| Model Number |
0279350201 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the device broke during the procedure.All pieces were retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: the serfas probe broke off in the joint during surgery.The patient had to be x-rayed to find the broken off pieces.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be excessive force used when inserting or removing the probe, a probe inserted into an obstructed passageway, or any forceful impact to the tip of the probe with rigid objects.Non-conforming component, poor assembly process, misuse.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that the device broke during the procedure.All pieces were retrieved.
|
| |
|