| Model Number |
UM-3R-3 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
An olympus employee reported on behalf of the customer, the ultrasonic probe ultrasound images are unclear and difficult to see during surgery.The frequency is not high.The procedure was completed using another device.There was no report of patient harm associated with this event.During incoming inspection, it was determined there was leakage in the ultrasonic medium due to perforation in the tip sheath.This medwatch is being submitted to capture the reportable malfunction found during evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
The subject device was returned to olympus for evaluation.During inspection and testing, the allegation of unclear images was not confirmed.In addition, there was leakage in the ultrasonic medium due to perforation in the tip sheath.Additional information has been requested regarding this event.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if additional information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely the ultrasonic image could not be rendered normally because the ultrasonic wave could not be transmitted normally due to the leakage of the ultrasonic medium inside and the contamination of air bubbles due to the occurrence of a hole in the sheath at the tip due to external force being applied.The event can be detected/prevented by following the instructions for use which state: "5.2.3 check for air bubbles - the ultrasound probe contains an ultrasound medium (liquid).Due to storage conditions, air bubbles may be generated around the ultrasonic probe or there may be no medium at the tip.Before using the ultrasound probe, be sure to check the air bubbles described here.If there is no medium or air bubbles around the ultrasonic probe, remove the air bubbles according to the instructions in "5.2.4 removal of the air bubbles".If there are bubbles on the ultrasonic transducer, a good ultrasonic image cannot be obtained." "5.2.2 inspection of appearance - inspection method: 1.Lightly pinch the probe insert with your finger and check that there are no noticeable breaks or media leaks over the entire length.2.Check the probe connection tube for dirt or foreign matter." olympus will continue to monitor field performance for this device.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated field: b5 and h10.This report is being supplemented to provide additional information based on the device evaluation.
|
| |
|
Event or Problem Description
|
|
Additional information was provided regarding this event.The intended procedure was a diagnostic ultrasound endoscopy.There was no delay in procedure.
|
| |
|