ABBOTT MEDICAL LAMITRODE 4 LEAD KIT, 60CM LENGTH; SCS PADDLE LEAD Back to Search Results
Model Number 3240
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code Inadequate Pain Relief (2388)
Date of Event 11/07/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event is estimated.Additional components potentially involved in the event include: common device name: scs lead, model: 3240, udi: (b)(4), serial:n/a, batch:5636036.
 
Event or Problem Description
It was reported that the patient experienced ineffective therapy due to high impedance on one of the occipital leads.As such, surgical intervention may take place at a later date to address the issue.Investigation was unable to determine which of the leads attributed to the event.
 
Event or Problem Description
Additional information received indicates that surgical intervention took place where in the left lead was explanted and replaced to address the issue.Reportedly, therapy was confirmed post op.
 
Additional Manufacturer Narrative
The reported event of ineffective stimulation was confirmed.As received, the lamitrode lead had all three internal wires broken, that would cause high impedance and is consistent with an overstress condition or sudden event the lead was subjected while in vivo.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAMITRODE 4 LEAD KIT, 60CM LENGTH
Common Device Name
SCS PADDLE LEAD
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key15933752
Report Number1627487-2022-06687
Device Sequence Number20473675
Product Code LGW
UDI-Device Identifier05414734402033
UDI-Public05414734402033
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P010032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/12/2018
Device Model Number3240
Device Catalogue Number3240
Device Lot Number5636036
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/15/2022
Supplement Date Received by Manufacturer12/16/2022
02/23/2023
Initial Report FDA Received Date12/07/2022
Supplement Report FDA Received Date01/10/2023
03/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/12/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SCS EXTENSION (X2).; SCS IPG.; SCS LEAD (X3).
Outcome Attributed to Adverse Event Other;
Patient Age34 YR
Patient SexFemale
Patient Weight59 KG
-
-