| Model Number |
IGF-1 |
| Medical Device Problem Codes |
Product Quality Problem (1506); Incorrect, Inadequate or Imprecise Result or Readings (1535); Improper or Incorrect Procedure or Method (2017); High Test Results (2457)
|
| Health Effect - Clinical Codes |
Hormonal Imbalance (4495); Insufficient Information (4580)
|
| Date of Event |
05/13/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
On follow-up, the reporter stated that it appears the assay is working as expected.The reporter stated that their physicians' concern was with the interpretation and comparison of the roche results to a previous method.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
We received an allegation that the elecsys igfbp-3 assay and elecsys igf-1 assay results may have contributed to a negative impact on patients' growth and the patients were reportedly not progressing as expected.This mdr is for the igf-1 assay.Refer to patient identifier: (b)(6) for the igfbp-3 assay.The reporter stated that pediatric endocrinologists have raised a concern about the igf-1 and igfbp3 results.They reportedly noticed that the values have been higher than the previous method (the immulite) and have been reportedly treating their patients based on the higher results.The reporter stated that during validation prior to changing to the roche method, the e411 was reportedly running higher than the immulite.Most samples reportedly stayed within or outside of the reference range.The reporter stated that they thought that the previous system was too low, not that the roche was too high.The reporter stated they had changed the reference range to reflect the roche reference range.This reportedly had increased slightly from the previous methodology.The reporter went on to state that after the physicians treated their patients based on roche's results, they saw a negative impact on their patients' growth and the patients were reportedly not progressing as expected.Additional information was requested, but could not be obtained despite multiple follow-up attempts.This information includes: clarification regarding the allegation: "a negative impact to their growth" and "patients were not progressing in growth as expected." clarification whether the therapeutic dosage changed for the affected children, or if the therapeutic dosage was scheduled from the beginning at an insufficient level? this mdr is being submitted with an abundance of caution.
|
| |
|
Additional Manufacturer Narrative
|
|
The serial number of the cobas e411 disk is (b)(6).Refer to attachment (b)(6) for the questionable results provided on 19-dec-2022.Medwatch field b1 was updated.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer confirmed their rejection of the allegation of unanticipated serious deterioration in health and that the statement that the patients were not progressing in growth as expected is no longer part of the allegation.At no point has anyone connected with the case provided any evidence of no progression in the growth of the pediatric patients as expected, despite having been asked multiple times on different days at different times.The calibration data, qc recovery, preanalytical data, alarm trace, and service information were requested but were not provided.The cause of the event could not be determined.
|
| |
|
Additional Manufacturer Narrative
|
|
The field application specialist (fas) collected calibration performance and other data for review.The last calibration performed on (b)(6) 2023 had calibrator signals that were slightly higher than the expected calibrator signals but this does not appear to affect the performance of the calibration.There were no alarms noted.The qc recovery on the date of the event was within the customer¿s established ranges.There is no indication of a performance issue with the reagent or instrument.Based on the data provided, the cause of the event could not be determined.A general instrument or reagent problem could not be identified.
|
| |
|
Search Alerts/Recalls
|