| Model Number |
3288 |
| Medical Device Problem Codes |
Material Fragmentation (1261); Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/07/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Unique device identifier (udi #): the udi is unknown because the lot number was not provided.
|
| |
|
Event or Problem Description
|
|
Related manufacturer reference number: 1627487-2022-06689.It was reported that during an ipg explant procedure the physician cut the leads.The cut portion of the leads remain implanted in the patient.Surgical intervention may be pending.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported issue for ¿infection¿ cannot be confirmed through product analysis or testing.Analysis of lead a found it was incomplete, only the terminal end segment was returned.Continuity was checked and no opens were found, setscrew engagement marks were observed.Images were taken to record as received condition.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported issue for ¿infection¿ cannot be confirmed through product analysis or testing.Analysis of lead a found it was incomplete, only the terminal end segment was returned.Continuity was checked and no opens were found, setscrew engagement marks were observed.Images were taken to record as received condition.
|
| |
|