| Model Number |
1012451-15 |
| Medical Device Problem Codes |
Material Rupture (1546); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/09/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat the proximal right coronary artery rca with moderate calcification and mild tortuosity.Pre-dilatation was performed with a non-abbott balloon and a non-abbott stent was implanted.For post-dilatation the 3.5x15 mm nc trek balloon dilatation catheter bdc was advanced with resistance with the anatomy noted and inflated but it ruptured during the first inflation at 18 atmospheres.A non-abbott balloon was used to complete the procedure.There were no adverse patient effects and there was no reported clinically significant delay in the procedure.No additional information was provided.
|
| |
|