| Model Number |
Z315H |
| Medical Device Problem Codes |
Break (1069); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Product not returned.Additional information has been requested and received.Attempts to obtain the device have been made, however not received to date.Please provide the lot number: lot number is scmclx.Device return status: we regularly contact with sales rep about the device returning.Additional information provided: on (b)(6) 2022, the subcutaneous dermal layer was buried sutured.After passing the needle through, when the tip of the needle was pinched with forceps and pulled, the suture and needle were cut off.There were 2 complaint products from the same box.The surgeon asked to investigate the remaining same lots for reference.The event occurred at (b)(6) 2022.Additional information has been requested however not received.Attempts to obtain the device have been made, however not received to date.If further details are received at a later date a supplemental medwatch will be sent.Please clarify the event: did the suture break per the event description (during a cesarean section, the suture broke during use) or was the needle broken as in the response: (on (b)(6) 2022, the subcutaneous dermal layer was buried sutured.After passing the needle through, when the tip of the needle was pinched with forceps and pulled, the suture and needle were cut off.There were 2 complaint products from the same box.The surgeon asked to investigate the remaining same lots for reference.) 2 events were mentioned.Is there another file to capture the 2nd event? or did this occur with this same patient with a 2nd suture? this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported a patient underwent a cesarean section on (b)(6) 2022 and a suture was used.The suture broke during use.Further details are not provided.There were no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 12/28/2022.H6 component code: g07002 no device problem found.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection evaluation was conducted on the returned sample.Visual analysis of the returned sample revealed that one empty labeled winding former and one suture piece of product code z315h were returned for analysis.During visual inspection of the sample, the suture piece was examined, and the ends of the strand were noted to be cut probably caused by a surgical instrument.In addition, body fluids were noted along the strands.The product code contains an absorbable suture and the time of exposure that the suture in the environment could not be determined; therefore, the functional test cannot be performed.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Additional h3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that one opened box with eight packets that pertain to product code z315h was returned with the packaging closed.Upon initial inspection, of the samples, no external damages were observed on the packets.In order to evaluate the conditions of the returned samples, the packets were opened, and no defects were detected.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related to damaged or breakage sutures and no defects were observed during the evaluation.A functional test was performed using instron equipment and the tensile strength force was above the minimum requirement.The event described could not be confirmed as the device performed without any defect noted.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Additional information was requested, the following was obtained: did the suture break per the event description (during a cesarean section, the suture broke during use) or was the needle broken as in the response: (on november 15, the subcutaneous dermal layer was buried sutured.After passing the needle through, when the tip of the needle was pinched with forceps and pulled, the suture and needle were cut off.There were 2 complaint products from the same box.The surgeon asked to investigate the remaining same lots for reference.) the suture was detached from needle.2 events were mentioned.Is there another file to capture the 2nd event? or did this occur with this same patient with a 2nd suture? (b)(4).The following information was requested: according to the additional information received for (b)(4), pull off suture needle occurred.Please clarify which event occurred for (b)(4): breakage suture or pull off suture needle? this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Date sent to the fda: 1/13/2023.Additional information was requested, the following was obtained: "according to the additional information received for (b)(4), pull off suture needle occurred.Please clarify which event occurred for (b)(4).Breakage suture or pull off suture needle? the suture was broken and cut from the needle.No further information will be provided." this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|