| Model Number |
CI-1400-01 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Intermittent Communication Failure (4038)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
11/10/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Event or Problem Description
|
|
The recipient is reportedly experiencing loss of lock.Device testing could not be completed due to loss of lock.Revision surgery will be scheduled.
|
| |
|
Additional Manufacturer Narrative
|
|
Advanced bionics considers the investigation into this reportable event as closed.The recipient will reportedly not undergo revision surgery at this time.A review of the device history record was completed and no anomalies were noted.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Additional Manufacturer Narrative
|
|
The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Additional Manufacturer Narrative
|
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed sliced silicone overmold on the bottom cover, which is believed to have occurred during revision surgery.Visual inspection also revealed broken antenna wires.The photographic imaging inspection confirmed broken antenna coil wires.System lock could not be obtained at any spacing.The no lock condition prevented an electrical test from being performed.The device passed some of the electrical tests performed.The device passed the mechanical tests performed.This device had broken antenna coil wires.A corrective action was implemented.This version of the hires 90k device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: section h.4.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Additional Manufacturer Narrative
|
|
Disclaimer: advanced bionics does not intend, that this report be any admission of liability, fault or product defect.
|
| |
|