| Model Number |
CI-1400-01 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Intermittent Communication Failure (4038)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
11/10/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The recipient is reportedly experiencing loss of lock.Device testing could not be completed due to loss of lock.Revision surgery will be scheduled.
|
| |
|
Additional Manufacturer Narrative
|
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Additional Manufacturer Narrative
|
|
Advanced bionics considers the investigation into this reportable event as closed.This is considered a failure in-situ.A review of the device history record was completed and no anomalies were noted.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Additional Manufacturer Narrative
|
|
Advanced bionics considers the investigation into this reportable event as closed.The recipient will reportedly not pursue revision surgery at this time.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Additional Manufacturer Narrative
|
|
The recipient's device was explanted.During revision surgery, the surgeon found ossification and was unable to insert the electrode the recipient was not reimplanted.The recipient has healed.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: section d.9 disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Additional Manufacturer Narrative
|
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed broken shield and antenna coil wires.Photographic imaging inspection confirmed broken shield and antenna coil wires.System lock could not be obtained at any spacing.The no lock condition prevented an electrical test from being performed.The device passed the electrical tests performed.The device passed the mechanical test performed.This device had broken antenna coil wires.A corrective action was implemented.This older device configuration is not currently manufactured.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|