• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES 90K¿ IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES 90K¿ IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1400-01
Medical Device Problem Codes Mechanical Problem (1384); Intermittent Communication Failure (4038)
Health Effect - Clinical Codes Insufficient Information (4580); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 11/10/2023
Type of Reportable Event Malfunction
Event or Problem Description
The recipient is reportedly experiencing loss of lock.Device testing could not be completed due to loss of lock.Revision surgery will be scheduled.
 
Additional Manufacturer Narrative
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Additional Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.This is considered a failure in-situ.A review of the device history record was completed and no anomalies were noted.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Additional Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The recipient will reportedly not pursue revision surgery at this time.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Additional Manufacturer Narrative
The recipient's device was explanted.During revision surgery, the surgeon found ossification and was unable to insert the electrode the recipient was not reimplanted.The recipient has healed.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Additional Manufacturer Narrative
Additional information: section d.9 disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Additional Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed broken shield and antenna coil wires.Photographic imaging inspection confirmed broken shield and antenna coil wires.System lock could not be obtained at any spacing.The no lock condition prevented an electrical test from being performed.The device passed the electrical tests performed.The device passed the mechanical test performed.This device had broken antenna coil wires.A corrective action was implemented.This older device configuration is not currently manufactured.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K¿ IMPLANT
Common Device Name
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
heidi white
28515 westinghouse place
valencia, CA 91355
MDR Report Key15936241
Report Number3006556115-2022-02139
Device Sequence Number6442009
Product Code MCM
UDI-Device Identifier07630016816253
UDI-Public(01)07630016816253(11)151208(17)171031
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P960058
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/31/2017
Device Model NumberCI-1400-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Not provided
Not provided
Not provided
Not provided
Initial Report FDA Received Date12/07/2022
Supplement Report FDA Received Date06/06/2023
06/29/2023
12/06/2023
05/13/2024
06/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/10/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age8 YR
Patient SexMale
-
-