It was reported that on (b)(6) 2022, a 27mm portico valve was chosen for implant using a large flexnav delivery system.The valve was sized as 77.4mm.During the procedure, valve was implanted at the recommended depth of 3mm on the non-coronary cusp.However, the operator misjudged the implantation process and the released the valve before having full annular contact.The valve reportedly migrated to -2mm, in the sinus.The valve was snared into the ascending aorta to stabilize it before performing a valve in valve procedure.An additional 27mm portico valve (19351309) was then implanted at 5mm.A mild leak was detected, and the gradient was measured at 4 mmhg.A post-dilatation was performed because of the leak.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in the procedure.The patient was reported as discharged.
|
An event of valve migration after deployment was reported.A returned device assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications.Possible contributing factors for valve migration are intra-procedural difficulties, implant depth, annular calcium distribution and deployment or retraction difficulties.Information from field indicated that there was sufficient calcium to allow anchoring of the device.However, the operator misjudged the implantation process and released the valve before having full annular contact causing the valve to migrate.Additional information indicated that the valve was snared and a valve in valve procedure was performed.Based on the information received, the reported incident is consistent with the valve being released in an undesirable position.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.Please note that per the instructions for use, "once the valve is fully deployed, repositioning and retrieval of the valve is not possible.Attempted retrieval (e.G., use of a guidewire, snare, or forceps) may cause aortic root, coronary artery, and/or myocardial damage.".
|